Global Global Algérie Français Algeria English Angola Português Angola English Argentina Español Argentina English Australia English Austria Deutsch Austria English Azerbaijan English Azerbaijan Русский Bahrain English Bangladesh English Belgium English Belgium Français Belgium Nederlands Brazil Português Brazil English Bulgaria български Bulgaria English Cameroon English Cameroon Français Canada English Canada Français Chile Español Chile English China 中文 China 日本語 China English Colombia Español Colombia English Croatia English Czech Republic Čeština Czech Republic English Denmark English Djibouti English Ecuador Español Ecuador English Egypt English Estonia English Ethiopia English Finland Suomi Finland English France Français France English Georgia English Georgia Русский Germany Deutsch Germany English Ghana English Greece Ελληνικά Greece English Guatemala Español Guatemala English Hong Kong, China English Hong Kong, China 中文 Hungary Magyar Hungary English India English Indonesia English Iraq English Ireland English Italy Italiano Italy English Ivory Coast Français Ivory Coast English Japan 日本語 Japan English Jordan English Kazakhstan Русский Kazakhstan English Kenya English Kuwait English Kyrgyzstan Русский Kyrgyzstan English Latvia English Lithuania English Malaysia English Mauritius English Mauritius français (Maurice) Mexico Español Mexico English Moldova Română Moldova English Mongolia English Morocco English Morocco Français Mozambique Português Mozambique English Netherlands Nederlands Netherlands English New Zealand English Nigeria English Norway English Oman English Pakistan English Paraguay Español Paraguay English Peru Español Peru English Philippines English Poland Polski Poland English Portugal Português Portugal English Qatar English Republic of Korea 한국어 Republic of Korea English Romania Română Romania English Saudi Arabia English Serbia Српски Serbia English Singapore English Slovakia English Slovenia English South Africa English Spain English Spain Español Sri Lanka English Sweden English Switzerland Deutsch Switzerland Français Switzerland Italiano Switzerland English Tanzania English Thailand ไทย Thailand English Togo English Togo Français Tunisia English Tunisia Français Türkiye Türkçe Türkiye English Turkmenistan Русский Turkmenistan English Ukraine Українська Ukraine English United Arab Emirates English United Kingdom English Uruguay Español Uruguay English USA English Uzbekistan English Uzbekistan Русский Vietnam Tiếng Việt Vietnam English

Oxygen permeability testing, commonly reported as the oxygen transmission rate (OTR), is a critical quality control and material qualification measure for evaluating the barrier performance of packaging used in food, pharmaceutical, and medical applications. This technical article explains the underlying principles, applicable international standards such as ASTM D3985, ISO 15105, and GB/T 19789, the types of samples that can be tested, and how results are interpreted and applied. It serves as a practical reference for packaging engineers, quality managers, and laboratory professionals seeking reliable oxygen barrier characterization.

Technical Overview

Oxygen transmission rate (OTR) quantifies the steady-state rate at which oxygen permeates through a packaging material under specified conditions of temperature and relative humidity. For flat materials such as films and laminates, OTR is typically expressed in cm³/(m²·24h), while finished containers are reported in cm³/(package·24h), both normalized to a partial pressure difference of one atmosphere. Closely related parameters include oxygen permeance, which accounts for the pressure driving force, and the oxygen permeability coefficient, which further normalizes for material thickness and is an intrinsic property of the polymer.

Oxygen ingress is one of the principal causes of product deterioration in packaged goods. In foods, it drives lipid oxidation, vitamin loss, discoloration, stale flavor development, and aerobic microbial growth. In pharmaceuticals and medical devices, oxygen can degrade active ingredients, reduce sterility assurance, and shorten shelf life. Accurate determination of oxygen permeability therefore underpins material selection, packaging design, shelf-life modeling, supplier verification, and regulatory compliance.

Barrier performance is strongly dependent on temperature and, for many materials, on relative humidity. Hydrophilic barrier polymers such as EVOH exhibit a significant loss of oxygen barrier properties at elevated humidity, while temperature increases accelerate permeation according to Arrhenius-type behavior. For this reason, oxygen permeability determination is performed under tightly controlled environmental conditions, most commonly 23°C at 0% RH or 50% RH, with additional testing at 38°C and higher humidity for accelerated or tropical climate scenarios.

Testing Samples

Flexible Films, Sheets, and Laminates

This category includes monolayer and coextruded films of PE, PP, PET, PA, and EVOH, metallized films, aluminum foil laminates, transparent oxide-coated (SiOx and AlOx) high-barrier films, and paper–plastic–aluminum composite structures. Flat specimens are die-cut to a defined test area and clamped between the two halves of a permeation cell.

Finished Packages and Rigid Containers

Complete packages such as bottles, jars, cans, trays, stand-up pouches, sachets, and blister packs can be tested as intact units. Whole-package testing captures the combined barrier contribution of the body material, seams, seals, and closures, and is essential for verifying real-world performance.

Closures, Liners, and Sealing Components

Cap liners, gaskets, induction seals, and closure materials may be evaluated individually as flat specimens or as part of a finished package assembly to quantify oxygen ingress pathways that bypass the primary barrier wall.

Coated, Treated, and High-Barrier Materials

Barrier coatings, plasma-deposited layers, and ultra-high-barrier laminates require instruments with low detection limits and long equilibration times. Testing these materials verifies coating integrity, uniformity, and durability after processing steps such as printing, lamination, or forming.

Testing Items

Typical test items and reported parameters for oxygen permeability determination include:

  • Oxygen transmission rate (OTR) of flat films, sheets, and laminates, in cm³/(m²·24h)
  • Oxygen transmission rate of finished packages and containers, in cm³/(package·24h)
  • Oxygen permeance, in cm³/(m²·24h·0.1MPa)
  • Oxygen permeability coefficient, in cm³·cm/(m²·24h·0.1MPa)
  • OTR as a function of relative humidity (for example, at 0%, 50%, 85%, and 90% RH)
  • OTR at elevated temperatures for accelerated or climatic studies
  • Comparative barrier ranking of candidate packaging structures
  • Shelf-life supporting data for oxygen-sensitive products

Testing Methods

Coulometric Sensor Method (Equal-Pressure Method)

The coulometric method is the most widely used technique for oxygen permeability determination and is specified in ASTM D3985, ISO 15105-2, GB/T 19789, and JIS K 7126-2. A conditioned specimen is clamped between two chambers: oxygen or air flows continuously across the outer side, while oxygen-free nitrogen carrier gas sweeps the inner side. Oxygen molecules that permeate the specimen are transported by the carrier gas to a coulometric sensor, where they are electrochemically reduced and generate a current directly proportional to the oxygen concentration. ASTM F1927 extends this technique to include precise relative humidity control on both sides of the specimen, which is essential for humidity-sensitive barrier materials.

Key advantages of the coulometric method include high sensitivity suitable for high-barrier materials, minimal mechanical stress on the specimen because both chambers are at essentially equal pressure, and excellent repeatability. The system is calibrated using certified reference films with known transmission rates.

Differential-Pressure (Manometric) Method

The differential-pressure method, covered by ASTM D1434, ISO 15105-1, and GB/T 1038, subjects one side of the specimen to oxygen at a known pressure while the other side is maintained at low pressure or evacuated. The increase in pressure on the low-pressure side, measured over time with a highly sensitive pressure transducer, is used to calculate the gas transmission rate and permeability coefficient. This method requires no carrier gas and can be adapted to other gases, although the applied pressure differential may deform very flexible films and sensitivity for ultra-high-barrier structures is generally lower than with the coulometric technique.

Whole-Package Testing

Finished packages are evaluated per ASTM F1307 using a coulometric detector. The sealed package is placed in a test chamber containing oxygen or air, while the package interior is purged with nitrogen through a needle arrangement. Oxygen permeating through the package walls, seals, and closure is carried to the sensor, yielding the total package oxygen ingress rate. This approach is indispensable because package-level barrier performance is frequently several times worse than that predicted from flat-film data alone due to seams, seals, and thinning during forming.

Test Conditions and Procedure

A typical determination proceeds as follows:

  • Specimen preparation and conditioning, usually for at least 48 hours at 23°C and 50% RH
  • Mounting of the specimen with a defined test area and verification of leak-tightness
  • Establishment of the specified test conditions, commonly 23°C at 0% or 50% RH
  • Purging of the system and calibration verification against a certified reference material
  • Monitoring of the sensor signal until steady-state transmission is reached
  • Calculation and reporting of OTR, permeance, or permeability with full test conditions

High-barrier materials may require several days to reach equilibrium, whereas conventional polyolefin films typically reach steady state within a few hours. All testing should be performed in accredited third-party laboratories or in controlled laboratory environments using calibrated instrumentation to ensure data integrity and comparability.

Applications

Oxygen permeability determination supports a broad range of industry applications:

  • Food and beverage packaging: qualification of films for modified atmosphere packaging (MAP), beer and juice bottles, edible oils, dairy products, roasted coffee, nuts, and dried foods where oxidation limits shelf life
  • Pharmaceutical packaging: evaluation of blister foils, laminate sachets, and bottle systems for oxidatively degradable tablets and capsules, and support of stability studies
  • Medical device packaging: verification of sterile barrier systems in accordance with ISO 11607 requirements and shelf-life justification
  • Cosmetics and personal care: barrier assessment for formulations sensitive to oxidation and rancidity
  • Research and development: screening of new barrier resins, coatings, and laminate structures, and optimization of down-gauging initiatives
  • Quality control: incoming material inspection, supplier comparison, and monitoring of process changes such as lamination, printing, or forming

FAQ

What is the difference between OTR, permeance, and permeability?

OTR is the raw transmission rate through a specific material or package under defined conditions. Permeance normalizes the OTR to the oxygen partial pressure difference across the specimen, and permeability further normalizes for thickness, making it an intrinsic material property useful for comparing different polymers.

Which test method should be selected?

The coulometric equal-pressure method is generally preferred for high-barrier films and for testing at controlled humidity, while the differential-pressure method is useful when multiple gases must be measured with a single instrument. Whole packages should always be tested by the package-level coulometric procedure to capture seal and closure effects.

Why must relative humidity be controlled during testing?

Many barrier polymers, most notably EVOH, are hygroscopic and lose a substantial portion of their barrier performance as humidity rises. Testing at the humidity representative of the product and distribution environment, or across a humidity range, provides realistic and comparable data.

How long does an oxygen permeability test take?

Standard polyolefin films typically reach steady state within several hours. High-barrier structures such as foil laminates, oxide-coated films, or thick EVOH coextrusions may require one to several days of equilibration before a valid steady-state value can be reported.

Why do package-level results differ from flat-film results?

Package-level oxygen ingress includes contributions from side seams, heat seals, closure interfaces, and material thinning during thermoforming or blow molding. These additional pathways commonly result in package OTR values higher than those extrapolated from flat-film measurements, which is why both levels of testing are recommended for critical applications.